Rohde & Schwarz ESIB: Difference between revisions
No edit summary |
No edit summary |
||
Line 11: | Line 11: | ||
The ESIB developement has come to a halt, the ESIB is replaced by the [[Rohde & Schwarz ESU]] | The ESIB developement has come to a halt, the ESIB is replaced by the [[Rohde & Schwarz ESU]] | ||
= RBW Filter settings = | |||
The ESIB has both 3 dB and 6 dB specific RBW filter settings. Some standards specify the use of 6dB RBW filters. | |||
If the ESIB is used in (scanning) receiver mode, the 6 dB RBW filters will always be used, because no 3 dB filters are implemented in the (scanning) receiver mode. | |||
As far as we know the 6 dB RBW filters in analyzer mode of the Rohde & Schwarz ESIB are only available in firmware version 2.07 and higher. If RadiMation detects that a firmware version of 2.07 and higher is used in the ESIB it will select the 6 dB filters in analyzer mode, otherwise it will use the 3 dB filters in analyzer mode. | |||
= Documentation = | = Documentation = |
Revision as of 07:10, 23 May 2007
The ESIB is a full compliant Spectrum analyzer/Receiver of Rohde & Schwarz.
The ESIB was formerly known as the ESI. Several products of Rohde & Schwarz are based on the ESIB receiver or are very similar:
- Rohde & Schwarz ESI
- Rohde & Schwarz ESIB
- Rohde & Schwarz ESPI
- Rohde & Schwarz ESCI
- Rohde & Schwarz FSP
- Rohde & Schwarz ESU
The ESIB developement has come to a halt, the ESIB is replaced by the Rohde & Schwarz ESU
RBW Filter settings[edit]
The ESIB has both 3 dB and 6 dB specific RBW filter settings. Some standards specify the use of 6dB RBW filters. If the ESIB is used in (scanning) receiver mode, the 6 dB RBW filters will always be used, because no 3 dB filters are implemented in the (scanning) receiver mode. As far as we know the 6 dB RBW filters in analyzer mode of the Rohde & Schwarz ESIB are only available in firmware version 2.07 and higher. If RadiMation detects that a firmware version of 2.07 and higher is used in the ESIB it will select the 6 dB filters in analyzer mode, otherwise it will use the 3 dB filters in analyzer mode.
Documentation[edit]
The programmer manual can be found here
Version history[edit]
- Firmware version 4.01 was released on 2001, juni 29
- Firmware version 4.33 was the latest version available in 2006, Februar