RadiMation Error: 805368369: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
No edit summary
 
(2 intermediate revisions by 2 users not shown)
Line 1: Line 1:
= A time out occured while waiting for the measurement result =
= A time out occurred while waiting for the measurement result =
A time out occured while waiting on a power measurement result of the U2004A
A time out occurred while waiting on a power measurement result of the U2004A. Depending on some specific settings, it can take longer than expected to perform a single power measurement. The single measurement is then taking more time than the time out, which causes this error.


== Solution ==
== Solution ==
* The measurement speed is related to the selected filter type. Please try different filter settings to increase the measurement speed. The filter settings can be found in the advanced settings of the device driver under the "Settings"-tab
Improve the measurement settings of the power meter in such a way that the measurement is faster.
* Increase the hidden MaxWaitTimeInSeconds property found in the Advanced-Settings of the device driver. This hidden setting can be accessed by pressing "shift+ Left Click" on the "Advanced" button of the device driver configuration.
* The measurement speed is related to the selected filter type. Use a lower filter setting to increase the measurement speed. The filter settings can be found in the advanced settings of the device driver under the {{ScreenElement|Settings}}-tab. For more details see the driver configuration of the [[Agilent Technologies U2000A]]
The hidden setting is found under the "[U2004A]" section
* Increase the hidden MaxWaitTimeInSeconds property found in the 'Advanced' settings of the device driver. This hidden setting can be accessed by pressing "shift+ Left Click" on the {{ScreenElement|Advanced}} button of the device driver configuration. The hidden setting is found under the "[U2004A]" section


[[Category:RadiMation Error]]
[[Category:RadiMation Error]]
[[Category:RadiMation]]
[[Category:RadiMation]]

Latest revision as of 10:27, 19 April 2023

A time out occurred while waiting for the measurement result

A time out occurred while waiting on a power measurement result of the U2004A. Depending on some specific settings, it can take longer than expected to perform a single power measurement. The single measurement is then taking more time than the time out, which causes this error.

Solution

Improve the measurement settings of the power meter in such a way that the measurement is faster.

  • The measurement speed is related to the selected filter type. Use a lower filter setting to increase the measurement speed. The filter settings can be found in the advanced settings of the device driver under the Settings-tab. For more details see the driver configuration of the Agilent Technologies U2000A
  • Increase the hidden MaxWaitTimeInSeconds property found in the 'Advanced' settings of the device driver. This hidden setting can be accessed by pressing "shift+ Left Click" on the Advanced button of the device driver configuration. The hidden setting is found under the "[U2004A]" section