Hameg communication: Difference between revisions
m (Jeru moved page Hameg to Hameg communication) |
|||
(2 intermediate revisions by one other user not shown) | |||
Line 1: | Line 1: | ||
== | == Connection methods == | ||
=== USB === | |||
The Hameg signal generator can be controlled by {{RadiMation}} by USB connection. This does require the installation of a specific windows driver provided by [[Rohde & Schwarz]]. | |||
Note the following when setting up to control the {{PAGENAME}} by USB: | |||
Note the following when setting up to control the | |||
The device comes available as a separate 'HAMEG Measurement Devices' category in the Windows Device manager. | The device comes available as a separate 'HAMEG Measurement Devices' category in the Windows Device manager. | ||
Line 15: | Line 9: | ||
Then reconnect the USB again, and load the additional USB drivers. | Then reconnect the USB again, and load the additional USB drivers. | ||
==== RS-232 | ==== Windows 11 support ==== | ||
The Windows USB driver provided by [[Rohde & Schwarz]] are not supported on Microsoft Windows 11 and newer. As an alternative the driver can be controlled using the hardware RS232 communication port of the {{PAGENAME}}. | |||
=== RS-232 === | |||
The following communication settings can be used: | The following communication settings can be used: | ||
Latest revision as of 14:03, 29 January 2024
Connection methods[edit]
USB[edit]
The Hameg signal generator can be controlled by RadiMation® by USB connection. This does require the installation of a specific windows driver provided by Rohde & Schwarz.
Note the following when setting up to control the Hameg communication by USB:
The device comes available as a separate 'HAMEG Measurement Devices' category in the Windows Device manager. It is necessary to enable the 'Load VCP' option of the driver in the Windows Device manager. Then reconnect the USB again, and load the additional USB drivers.
Windows 11 support[edit]
The Windows USB driver provided by Rohde & Schwarz are not supported on Microsoft Windows 11 and newer. As an alternative the driver can be controlled using the hardware RS232 communication port of the Hameg communication.
RS-232[edit]
The following communication settings can be used:
Baudrate: 4800
Data bits: 8
Parity: None
Stop bits: 1