RadiMation Error: 805307458

From RadiWiki
Revision as of 13:09, 8 September 2006 by Joro (talk | contribs)
Jump to navigation Jump to search

RadiSense fieldprobe not connected

A RadiSense field probe that is tried to be controlled is not connected. The controlled device returned a response ':E9', which means that the fieldprobe is not connected.

Solution

  • Check if the controlled device is powered on, and connected to the communication bus
  • Check if fieldprobe is connected
  • Check if the correct settings for the communication bus are specified

For RadiSense probes, which are powered by LASER light:

  • Check if the probe is 'activated', that the LASER is turned on.