= There is no average probe selected =
 
= There is no average probe selected =
 +
When a Radiated Immunity [[Chapter_12#One-point_calibration  1 point calibration]] or a Radiated Immunity Fixed Field test is done, which used the measurement value of the field probe to regulate the test level, the regulation is done on the average of the field probes that are measured. It is thus required that at least one electrical field strength probe is selected to be included in the calculation of the average field.
  
This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected.
+
This error happens when you are doing a Radiated Immunity calibration or a Radiated Immunity Fixed Field and do not have at least one averaging probe selected.  
For a one-point radiated immunity calibration at least one electrical field strength probe should be selected to be included in the calculation of the average field.
 
  
 
Also see [[Chapter_12#One-point_calibration|    One-point calibration]].
 
Also see [[Chapter_12#One-point_calibration|    One-point calibration]].

Revision as of 09:29, 19 January 2022

There is no average probe selected

When a Radiated Immunity Chapter_12#One-point_calibration 1 point calibration or a Radiated Immunity Fixed Field test is done, which used the measurement value of the field probe to regulate the test level, the regulation is done on the average of the field probes that are measured. It is thus required that at least one electrical field strength probe is selected to be included in the calculation of the average field.

This error happens when you are doing a Radiated Immunity calibration or a Radiated Immunity Fixed Field and do not have at least one averaging probe selected.

Also see One-point calibration.

Solution

Re-open your Radiated Immunity calibration test file and go to 'Field Probes' and make sure that at least one of the probes has the averaging checkbox checked.