Difference between revisions of "RadiMation Error: 536871966"

From RadiWiki
Jump to: navigation, search
(There is no average probe selected)
Line 2: Line 2:
  
 
This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected.  
 
This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected.  
For a RI calibration at least one averaging probe is mandatory to do the calibration.  
+
For a one-point radiated immunity calibration at least one electrical field strength probe should be selected to be included in the calculation of the average field.
 +
 
 +
Also see [[Chapter_12#One-point_calibration    One-point calibration]].
  
 
== Solution ==
 
== Solution ==

Revision as of 14:49, 19 May 2021

There is no average probe selected

This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a one-point radiated immunity calibration at least one electrical field strength probe should be selected to be included in the calculation of the average field.

Also see Chapter_12#One-point_calibration One-point calibration.

Solution

Re-open your Radiated Immunity calibration test file and go to 'Field Probes' and make sure that at least one of the probes has the averaging checkbox checked.