(Created page with "= Measuring at 150 kHz in a sweep with an activated "preselector" is not possible. Create a band below 150 kHz and a band above 150 kHz. = The R&S ESI analyser won't allow to...")
 
 
(One intermediate revision by the same user not shown)
Line 2: Line 2:
 
The R&S ESI analyser won't allow to set a start frequency below 150 kHz and end frequency at 150 kHz or higher with the preselector enabled.
 
The R&S ESI analyser won't allow to set a start frequency below 150 kHz and end frequency at 150 kHz or higher with the preselector enabled.
  
== Possible solutions ==
+
== Solution ==
 
* Create a multiband test with a band to max 149 kHz and one band from 150 kHz and higher.
 
* Create a multiband test with a band to max 149 kHz and one band from 150 kHz and higher.
* Wen not using a multiband test, then create two tests.
+
* When not using a multiband test, then create two tests. Where the first test stops at 150 kHz, and the second test starts at 150 kHz.
 +
 
 +
[[Category:RadiMation Error]]
 +
[[Category:RadiMation]]

Latest revision as of 08:24, 9 April 2021

Measuring at 150 kHz in a sweep with an activated "preselector" is not possible. Create a band below 150 kHz and a band above 150 kHz.

The R&S ESI analyser won't allow to set a start frequency below 150 kHz and end frequency at 150 kHz or higher with the preselector enabled.

Solution

  • Create a multiband test with a band to max 149 kHz and one band from 150 kHz and higher.
  • When not using a multiband test, then create two tests. Where the first test stops at 150 kHz, and the second test starts at 150 kHz.