GraphExport.CALFile.MinimumField

From RadiWiki
Jump to navigation Jump to search

This advanced option specifies the minimum electrical field that should at least be achieved to allow the export of the field strength graph to a radiated immunity calibration file.

If on one or more frequencies the field strength is below 1.0 V/m, the field strength cannot be exported to a radiated immunity graph, and the validation error is shown: "The electrical field graph contains data points that are below 1 V/m. The electrical field graph should at least be 1 V/m on all data points."

By default this value is 1.0 V/m. If for some reason it is required to allow export of an electrical field graph to a radiated immunity graph, to also contain values lower than 1.0 V/m, it is possible to set this value to 0.0 V/m, which will allow to export to the calibration file, even if the field strength was 0.0 V/m at one or more frequencies.

Warning-48px.png
Warning: Any change and/or modification to the advanced options, can result in unexpected results and/or incorrect measurement results. Changes in any advanced option should be verified and checked very carefully!