RadiMation Error: 536871960

From RadiWiki
Jump to navigation Jump to search

End Frequency is too high for the selected test-level file. <FILE>[edit]

This error indicates that a variable test-level file is used during an immunity test. However the end frequency of the test is higher than the highest frequency in the selected variable test-level file. The <FILE>-marker shows the exact name of the variable test-level file that was selected.

This error is most often triggered as a result of selecting the wrong variable test-level file, or specifying a too high end-frequency.

Solution[edit]

  • Decrease the end-frequency of the TSF file, to be equal or lower than the highest frequency in the selected variable test-level file.
  • Select another (correct) variable test-level file.
  • Update the variable test-level file, to include a higher end-frequency.