RadiMation Error: 805306941

From RadiWiki
Jump to: navigation, search

GPIB system driver error (EDVR): Ethernet communication problem

A RadiMation device driver is trying to control a GPIB device through a GPIB board. However an error is raised by the GPIB system driver that controls that GPIB board. The driver has detected that there is an ethernet communication problem with the GPIB Ethernet device.

National Instruments NI-488.2

This error corresponds to the National Instruments NI-488.2 error 0xxE1060075.

This error is returned when the driver is unable to communicate with a GPIB Ethernet interface during an ibfind or ibdev call for one of the following reasons:

  • The IP address is incorrect
  • The box is not powered on
  • The box did not pass its self-diagnostic test
  • In general, the box is not configured or installed correctly

Solution

When a National Instruments GPIB board is used, the configuration of the GPIB board can be checked and changed in the NI-MAX software.

  • Check if the GPIB Ethernet device is still present and available in the NI-MAX software.