RadiMation Error: 805307458
Jump to navigation
Jump to search
RadiSense fieldprobe not connected
A RadiSense field probe that is tried to be controlled is not connected. The controlled device returned a response ':E9', which means that the fieldprobe is not connected.
Solution
- Check if the controlled device is powered on, and connected to the communication bus
- Check if fieldprobe is connected
- Check if the correct settings for the communication bus are specified
For RadiSense probes, which are powered by laser light:
- Check if the probe is 'activated', that the laser is turned on.