RadiMation Error: 805364370
Jump to navigation
Jump to search
An invalid characterization is specified for the calibration kit name. Valid names are: '....'
If an automatic calibration unit is connected to the Rohde & Schwarz ZNB network analyser, the calibration kit name that is specified in RadiMation is used to select the characterization that is available for the automatic calibration unit. It is now detected that the name that is specified for the calibration kit name, is not a valid characterization name. The actual valid names are also listed in the error message.
Solution
- Specify a name for the calibration kit to use, as one of the available characterizations from the connected automatic calibration kit that is connected to the Rohde & Schwarz ZNB.
- The characterization with the name 'Factory' is always present, and can thus always be specified. Specifying an empty calibration kit name will also select this 'Factory' characterization.