Talk:RadiMation Application Note 100
Jump to navigation
Jump to search
MAHE: I have applied the requested changes
I do have a few remarks for the improvements of this application note:
- The last screenshot: (File:EnhancedWindowWithTest.png), shows an EUT window, a test and the Enhanced status window. The EUT Window is not relevant for the purpose of the screenshot, but it does cause that the screenshot is very wide. I think we should improve the screenshot to NOT show the EUT window.
- The values on all the screenshots that show the enhanced status window are very strange. They should be updated to show more logical values (maybe configure it in the KETC??):
- The last screenshot (File:EnhancedWindowWithTest.png) shows: -200 dBm signal power (How is that possible??) while the field is expected to be almost 100 V/m. That is practically not possible. Also the frequency is just too low. At 280 kHz no 100 V/m can be generated.
- The horizontal screenshot shows a frequency of 1 Hz.
- The vertical screenshot shows a frequency of 1 Hz
Review status[edit]
Applied review changes. Mahe (talk) 09:45, 12 Februari 2019 (CET)
Did review. See remarks above. Joro (talk) 09:39, 30 January 2019 (CET)