Difference between revisions of "RadiMation Error: 805366371"
(Created page with "The attenuator and the preamp can not be used at the same time.") |
(→Solution) |
||
(3 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
− | The attenuator and the | + | ==Cannot use the preamp and attenuator at the same time== |
+ | The AFJ receiver is controlled using a test configuration that has a fixed value for the attenuator and the preamplifier. This is not supported by the AFJ receiver itself. | ||
+ | |||
+ | ==Solution== | ||
+ | * Only use the attenuator or the preamplifier, the device does not support both settings at the same time. The other setting should be set to Auto | ||
+ | |||
+ | [[Category:RadiMation]] | ||
+ | [[Category:RadiMation Error]] |
Latest revision as of 08:05, 19 July 2024
Cannot use the preamp and attenuator at the same time
The AFJ receiver is controlled using a test configuration that has a fixed value for the attenuator and the preamplifier. This is not supported by the AFJ receiver itself.
Solution
- Only use the attenuator or the preamplifier, the device does not support both settings at the same time. The other setting should be set to Auto