The attenuator and the preamp can not be used at the same time.
==Cannot use the preamp and attenuator at the same time==
The AFJ receiver is controlled using a test configuration that has a fixed value for the attenuator and the preamplifier. This is not supported by the AFJ receiver itself.
 
==Solution==
* Only use the attenuator or the preamplifier, the device does not support both settings at the same time. The other setting should be set to Auto


[[Category:RadiMation]]
[[Category:RadiMation]]
[[Category:RadiMation Error]]
[[Category:RadiMation Error]]

Latest revision as of 08:05, 19 July 2024

Cannot use the preamp and attenuator at the same time

The AFJ receiver is controlled using a test configuration that has a fixed value for the attenuator and the preamplifier. This is not supported by the AFJ receiver itself.

Solution