== Attenuation / Gain System Calibration == | == Attenuation / Gain System Calibration == | ||
Before {{ | Before {{RadiMation}} can start to determine the attenuation or gain of a system, the system need to be calibrated. | ||
Because this type of test considered to be a calibration all eut windows must be closed. | Because this type of test considered to be a calibration all eut windows must be closed. | ||
{{menu|Calibration|System Calibration|Attenuation / Gain}} | {{menu|Calibration|System Calibration|Attenuation / Gain}} | ||
Line 6: | Line 6: | ||
[[Image:Attenuation System Calibration Configuration Window.png]] | [[Image:Attenuation System Calibration Configuration Window.png]] | ||
{{Frequency window immunity}} | {{Frequency window immunity}} | ||
{{ScreenElementDescriptionStart}} | {{ScreenElementDescriptionStart}} | ||
{{ScreenElementDescription|Signal power|The setting of the signal power level during the test.}} | {{ScreenElementDescription|Signal power|The setting of the signal power level during the test.}} | ||
{{ScreenElementDescription|Amount of samples|The amount of samples level during the test.}} | {{ScreenElementDescription|Amount of samples|The amount of samples level during the test.}} | ||
{{Test site window}} | {{Test site window}} | ||
{{ScreenElementDescription|Run Calibration|Start the configured calibration.}} | {{ScreenElementDescription|Run Calibration|Start the configured calibration.}} | ||
{{ScreenElementDescription|Cancel|Cancel the configured calibration.}} | {{ScreenElementDescription|Cancel|Cancel the configured calibration.}} | ||
{{ScreenElementEnvironmentButton}} | |||
{{ScreenElementNoteButton}} | {{ScreenElementNoteButton}} | ||
{{ScreenElementUnitsButton}} | |||
{{ScreenElementDescriptionEnd}} | {{ScreenElementDescriptionEnd}} | ||
Line 43: | Line 35: | ||
===Test set up=== | ===Test set up=== | ||
[[image:Attenuation Test Setup System Calibration.png]] | [[image:Attenuation Test Setup System Calibration.png]] | ||
{| | {|class="wikitable" | ||
!Device name || Tab in {{ | !# ||Device name || Tab in {{RadiMation}} configuration window | ||
|- | |- | ||
|Signal Generator || | | ||Signal Generator || devices 1 | ||
|- | |- | ||
|Power meter | | ||Forward Power meter || devices 1 | ||
|} | |} | ||
== Attenuation / Gain Test == | == Attenuation / Gain Test == | ||
Because the determination of gain or attenuation is considered to be a test on an actual EUT, {{ | Because the determination of gain or attenuation is considered to be a test on an actual EUT, {{RadiMation}} needs to have an EUT window open before this test can be performed. | ||
{{menu|Calibration| | {{menu|Calibration|EUT Calibration|Attenuation / Gain}} | ||
[[image:Attenuation EUT Calibration Configuration Window.png]] | [[image:Attenuation EUT Calibration Configuration Window.png]] | ||
{{ScreenElementDescriptionStart}} | |||
{{ScreenElementDescription|Description| }} | |||
{{ScreenElementDescriptionEnd}}{{Frequency window immunity}}{{ScreenElementDescriptionStart}} | |||
{{ScreenElementDescription|Signal power|The setting of the signal power level during the test.}} | |||
{{ScreenElementDescription|amount of samples|The amount of samples level during the test}} | |||
{{ScreenElementDescription|system Calibration|}} | |||
{{ScreenElementDescription|Test equipment|In the test site window the test engineer can select which equipment list will be used during this test.}} | |||
{{ScreenElementDescription|Test engineer|In the test engineer window, the test engineer can select its own name. The test engineer name will be stored by the test results}} | |||
{{ScreenElementDescription|Run Calibration|Start the configured calibration.}} | |||
{{ScreenElementDescription|Cancel|Cancel the configured calibration}} | |||
{{ScreenElementDescription|Limit lines|}} | |||
{{ScreenElementEnvironmentButton}} | |||
{{ScreenElementNoteButton}} | |||
{{ScreenElementUnitsButton}} | |||
{{ScreenElementDescription|Reporting|}} | |||
{{ScreenElementDescription|General Info|}} | |||
{{ScreenElementDescriptionEnd}} | |||
The only selectable option is the calibration file, when another calibration file is selected the settings of this file will be displayed. | The only selectable option is the calibration file, when another calibration file is selected the settings of this file will be displayed. | ||
Line 66: | Line 76: | ||
[[image:Attenuation EUT Calibration Result Window.png]] | [[image:Attenuation EUT Calibration Result Window.png]] | ||
{{ScreenElementDescriptionStart}} | {{ScreenElementDescriptionStart}} | ||
{{ScreenElementDescription|Settings| }} | |||
{{ScreenElementDescription|Attenuation| }} | |||
{{ScreenElementDescription|Dynamic range| }} | |||
{{ScreenElementDescription|Test equipment| }} | |||
{{ScreenElementDescription|Test engineer| }} | |||
{{ScreenElementDescription|Event-list| }} | |||
{{ScreenElementDescription|Ok|Close the current window.}} | |||
{{ScreenElementDescription|Create correction file|Create a correction file based on the test information.}} | {{ScreenElementDescription|Create correction file|Create a correction file based on the test information.}} | ||
{{ScreenElementDescription|Limit lines| }} | |||
{{ScreenElementEnvironmentButton}} | |||
{{ScreenElementNoteButton}} | {{ScreenElementNoteButton}} | ||
{{ScreenElementUnitsButton}} | |||
{{ScreenElementDescription|Reporting| }} | |||
{{ScreenElementDescription|General Info| }} | |||
{{ScreenElementDescriptionEnd}} | {{ScreenElementDescriptionEnd}} | ||
[[Category:RadiMation]] |
Latest revision as of 18:48, 6 October 2024
Attenuation / Gain System Calibration[edit]
Before RadiMation® can start to determine the attenuation or gain of a system, the system need to be calibrated. Because this type of test considered to be a calibration all eut windows must be closed.
- Calibration
- System Calibration
- Attenuation / Gain
- System Calibration
- Calibration
Frequency Range | Sets the frequency specification to a frequency range. |
Start | Sets start frequency. |
End | Sets end frequency. |
Step Fixed | Sets a fixed step size from start to end frequency. |
Step Previous Frequency | Sets a logarithmic step size from start to stop frequency |
Starting From | Sets the start frequency for a logarithmic step from which the actual measurement frequencies will be calculated. the starting from frequency should be lower then the actual start frequency. The test will start at the first calculated frequency that is equal or higher then the specified start frequency. |
Frequency list | With frequency list the operator can select a frequency list file by browsing to the file location. The frequencies inside the frequency list file will be measured. |
Signal power | The setting of the signal power level during the test. |
Amount of samples | The amount of samples level during the test. |
Test equipment | In the test site window the test engineer can select which equipment list will be used during this test. |
Test engineer | In the test engineer window, the test engineer can select its own name. The test engineer name will be stored by the test results. |
Equipment icon | By clicking on the equipment icon, the equipment list can be viewed and edited. |
Run Calibration | Start the configured calibration. |
Cancel | Cancel the configured calibration. |
Environment | Display the environment window. |
Note | Display the note window. |
Units | Display the Units configuration window. |
Running the test[edit]
After the test initialization has been performed and passed the user will be prompt.
Connect the power sensor to the power reference and press ok. The power sensor will now be calibrated using the power reference source to make sure the power sensor is as accurate as possible.
Note: | Not all power sensors have a power reference source, in this case leave the power sensor connected to the calibration set up. |
Connect the power sensor to the calibration set up and press ok.
After the calibration has been completed the user will be prompt where the calibration file needs to be saved.
Test set up[edit]
# | Device name | Tab in RadiMation® configuration window |
---|---|---|
Signal Generator | devices 1 | |
Forward Power meter | devices 1 |
Attenuation / Gain Test[edit]
Because the determination of gain or attenuation is considered to be a test on an actual EUT, RadiMation® needs to have an EUT window open before this test can be performed.
- Calibration
- EUT Calibration
- Attenuation / Gain
- EUT Calibration
- Calibration
Description |
Frequency Range | Sets the frequency specification to a frequency range. |
Start | Sets start frequency. |
End | Sets end frequency. |
Step Fixed | Sets a fixed step size from start to end frequency. |
Step Previous Frequency | Sets a logarithmic step size from start to stop frequency |
Starting From | Sets the start frequency for a logarithmic step from which the actual measurement frequencies will be calculated. the starting from frequency should be lower then the actual start frequency. The test will start at the first calculated frequency that is equal or higher then the specified start frequency. |
Frequency list | With frequency list the operator can select a frequency list file by browsing to the file location. The frequencies inside the frequency list file will be measured. |
Signal power | The setting of the signal power level during the test. |
amount of samples | The amount of samples level during the test |
system Calibration |
Test equipment | In the test site window the test engineer can select which equipment list will be used during this test. |
Test engineer | In the test engineer window, the test engineer can select its own name. The test engineer name will be stored by the test results |
Run Calibration | Start the configured calibration. |
Cancel | Cancel the configured calibration |
Limit lines |
Environment | Display the environment window. |
Note | Display the note window. |
Units | Display the Units configuration window. |
Reporting |
General Info |
The only selectable option is the calibration file, when another calibration file is selected the settings of this file will be displayed.
Connect the test set up and press Run Calibration.
Once the actual test has been performed the test result file can be opened in the EUT window.
Settings |
Attenuation |
Dynamic range |
Test equipment |
Test engineer |
Event-list |
Ok | Close the current window. |
Create correction file | Create a correction file based on the test information. |
Limit lines |
Environment | Display the environment window. |
Note | Display the note window. |
Units | Display the Units configuration window. |
Reporting |
General Info |