RadiMation Error: 536871966: Difference between revisions
Jump to navigation
Jump to search
(Created page with "= There is no average probe selected = This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a RI...") |
|||
Line 5: | Line 5: | ||
== Solution == | == Solution == | ||
Re-open your Radiated Immunity calibration test file and go to | Re-open your Radiated Immunity calibration test file and go to 'Field Probe' and make sure that at least one of the probes has the averaging checkbox checked. | ||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] | ||
[[Category:RadiMation]] | [[Category:RadiMation]] |
Revision as of 12:34, 6 June 2019
There is no average probe selected[edit]
This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a RI calibration at least one averaging probe is mandatory to do the calibration.
Solution[edit]
Re-open your Radiated Immunity calibration test file and go to 'Field Probe' and make sure that at least one of the probes has the averaging checkbox checked.