Teseq ITS 6006: Difference between revisions
DevicesBot (talk | contribs) (Updated DeviceDriverInfo template) |
DevicesBot (talk | contribs) (Updated DeviceDriverInfo template) |
||
Line 2: | Line 2: | ||
|DLLFile=RADCOMSG.DLL | |DLLFile=RADCOMSG.DLL | ||
|DeviceBrand=Teseq | |DeviceBrand=Teseq | ||
|DeviceDriverName= | |DeviceDriverName=TESEQ_ITS6006_SG | ||
|DeviceType=ITS 6006 | |DeviceType=ITS 6006 | ||
|HideAutoData=1 | |HideAutoData=1 | ||
|SupportedDeviceTypes= | |SupportedDeviceTypes=1 | ||
}} | }} | ||
Revision as of 23:50, 16 March 2020
The Teseq ITS 6006 device driver is a Signal Generator which is supported by RadiMation®.
Configuration[edit]
The following tabs are available in the advanced configuration of the Teseq ITS 6006:
The frequency range of the Teseq ITS 6006 as provided by the manufacturer is shown and selected as default. It is possible to overrule these frequencies and to manual adjust the allowed frequency range of the Teseq ITS 6006.
Use default start frequency | If the checkbox is checked, the default start frequency will be used as the lowest usable frequency in a test for this device. |
Custom start frequency | If the Use default start frequency checkbox is unchecked, another start frequency (expressed in MHz) can be specified. The customized start frequency will then be used as the lowest usable frequency in a test for this device. The customized frequency can be a limitation or an extension of the default start frequency. |
Use default stop frequency | If the checkbox is checked, the default stop frequency will be used as the highest usable frequency in a test for this device. |
Custom stop frequency | If the Use default stop frequency checkbox is unchecked, another stop frequency (expressed in MHz) can be specified. The customized stop frequency will then be used as the highest usable frequency in a test for this device. The customized frequency can be a limitation or an extension of the default stop frequency. |
Specifying a different frequency range can be useful if for example:
- A device (like a coupler, antenna, injection device, cable, etc...) is still useable (but out of specification) outside the standard suggested frequency range.
- An external mixer is used to measure an extended frequency range.
- An up- or down-convertor is used to shift the frequency range.
- A newer model of a device is present that has an extended frequency range, and still uses the same remote control commands.
Be careful changing these setting as RadiMation® is no longer able to verify if the Teseq ITS 6006 is used outside frequency range that is specified by the manufacturer. This may result to serious damage of your measurement device.
The USB drivers for the ITS-6006 are included in the 'ITS 6006 Tools Software (MS Windows)' and can be downloaded from: http://www.teseq.com/support/its-6006/its-6006-tools-software.php
When the RS232 connection is used, the communication settings should be set to: 115200 Baud, 8 bits, no parity, 1 stopbit. When the USB connection is used, the correct USB Virtual Comport should be selected, and the communication settings should be set to: 9600 baud, 8 bits, no parity, 1 stopbit.
These Teseq USB drivers are probably not working on Windows 10. It is advised to use Windows 7.