RadiMation Error: 80554641: Difference between revisions
Jump to navigation
Jump to search
(Created page with "=An attenuation other dan 20 dB is configured with the preamplifier enabled. Please disable the preamplifier or choose an attenuation of 20 dB= During the start of a new swee...") |
|||
Line 9: | Line 9: | ||
== Solution == | == Solution == | ||
* Disable the preamplifier or choose an attenuation of 20 dB. | * Disable the preamplifier or choose an attenuation of 20 dB. | ||
If this problem remains after manually checking the above options, further investigation by | If this problem remains after manually checking the above options, further investigation by Raditeq is necessary to find the real cause of this problem. | ||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] | ||
[[Category:RadiMation]] | [[Category:RadiMation]] |
Latest revision as of 08:48, 18 January 2021
An attenuation other dan 20 dB is configured with the preamplifier enabled. Please disable the preamplifier or choose an attenuation of 20 dB[edit]
During the start of a new sweep it is detected that the pre-amplifier is enabled while also an attenuation other than 20 dB is selected.
The device which returns this error shall set the attenuation to 20 when the preamplifier is enabled, doing so any other configured attenuation will be silently ignored.
Possible reasons for this error can be: the preamplifier is enabled while also an attenuation other dan 20 dB is selected.
Solution[edit]
- Disable the preamplifier or choose an attenuation of 20 dB.
If this problem remains after manually checking the above options, further investigation by Raditeq is necessary to find the real cause of this problem.