Difference between revisions of "RadiMation Error: 805306945"

From RadiWiki
Jump to: navigation, search
(Created page with "= GPIB system driver error (EDVR): Memory access violation = A RadiMation device driver is trying to control a GPIB device through a GPIB board. However an error is raised...")
(No difference)

Revision as of 17:48, 16 February 2021

GPIB system driver error (EDVR): Memory access violation

A RadiMation device driver is trying to control a GPIB device through a GPIB board. However an error is raised by the GPIB system driver that controls that GPIB board. The driver has detected that an illegal memory location was accessed.

Solution

When a National Instruments GPIB board is used, the configuration of the GPIB board can be checked and changed in the NI-MAX software.

  • Check if the GPIB device is still present and available in the NI-MAX software.
  • Disconnect and connect the GPIB device again.