RadiMation Error: 805366369: Difference between revisions
Jump to navigation
Jump to search
(Created page with "Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands: Band A: 9 kHz - 150 kHz (200 Hz) Band B: 150 kHz - 30 M...") |
No edit summary |
||
(2 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
==Frequency span with multiple RBW values setup== | |||
Band A: 9 kHz - 150 kHz (200 Hz) | ==Solution== | ||
Band B: 150 kHz - 30 MHz (9 kHz) | Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands or separate tests: | ||
Band C: 30 MHz - 1 GHz (120 kHz) | |||
* Band A: 9 kHz - 150 kHz (200 Hz) | |||
* Band B: 150 kHz - 30 MHz (9 kHz) | |||
* Band C: 30 MHz - 1 GHz (120 kHz) | |||
[[Category:RadiMation]] | |||
[[Category:RadiMation Error]] |
Latest revision as of 13:03, 13 April 2022
Frequency span with multiple RBW values setup[edit]
Solution[edit]
Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands or separate tests:
- Band A: 9 kHz - 150 kHz (200 Hz)
- Band B: 150 kHz - 30 MHz (9 kHz)
- Band C: 30 MHz - 1 GHz (120 kHz)