RadiMation Error: 805368369: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
No edit summary
Line 1: Line 1:
= A time out occured while waiting for the measurement result =
= A time out occurred while waiting for the measurement result =
A time out occured while waiting on a power measurement result of the U2004A
A time out occurred while waiting on a power measurement result of the U2004A. Depending on some specific settings, it can take longer than expected to perform a single power measurement.


== Solution ==
== Solution ==
* The measurement speed is related to the selected filter type. Please try different filter settings to increase the measurement speed. The filter settings can be found in the advanced settings of the device driver under the "Settings"-tab
Improve the settings of the power meter in such a way that the measurement is faster.
* Increase the hidden MaxWaitTimeInSeconds property found in the Advanced-Settings of the device driver. This hidden setting can be accessed by pressing "shift+ Left Click" on the "Advanced" button of the device driver configuration. The hidden setting is found under the "[U2004A]" section
* The measurement speed is related to the selected filter type. Use a lower filter setting to increase the measurement speed. The filter settings can be found in the advanced settings of the device driver under the {{ScreenElement|Settings}}-tab.
* Increase the hidden MaxWaitTimeInSeconds property found in the 'Advanced' settings of the device driver. This hidden setting can be accessed by pressing "shift+ Left Click" on the {{ScreenElement|Advanced}} button of the device driver configuration. The hidden setting is found under the "[U2004A]" section


[[Category:RadiMation Error]]
[[Category:RadiMation Error]]
[[Category:RadiMation]]
[[Category:RadiMation]]

Revision as of 11:30, 22 September 2022

A time out occurred while waiting for the measurement result

A time out occurred while waiting on a power measurement result of the U2004A. Depending on some specific settings, it can take longer than expected to perform a single power measurement.

Solution

Improve the settings of the power meter in such a way that the measurement is faster.

  • The measurement speed is related to the selected filter type. Use a lower filter setting to increase the measurement speed. The filter settings can be found in the advanced settings of the device driver under the Settings-tab.
  • Increase the hidden MaxWaitTimeInSeconds property found in the 'Advanced' settings of the device driver. This hidden setting can be accessed by pressing "shift+ Left Click" on the Advanced button of the device driver configuration. The hidden setting is found under the "[U2004A]" section