Line 1: | Line 1: | ||
= A time out | = A time out occurred while waiting for the measurement result = | ||
A time out | A time out occurred while waiting on a power measurement result of the U2004A. Depending on some specific settings, it can take longer than expected to perform a single power measurement. The single measurement is then taking more time than the time out, which causes this error. | ||
== Solution == | == Solution == | ||
* The measurement speed is related to the selected filter type. | Improve the measurement settings of the power meter in such a way that the measurement is faster. | ||
* Increase the hidden MaxWaitTimeInSeconds property found in the Advanced | * The measurement speed is related to the selected filter type. Use a lower filter setting to increase the measurement speed. The filter settings can be found in the advanced settings of the device driver under the {{ScreenElement|Settings}}-tab. For more details see the driver configuration of the [[Agilent Technologies U2000A]] | ||
* Increase the hidden MaxWaitTimeInSeconds property found in the 'Advanced' settings of the device driver. This hidden setting can be accessed by pressing "shift+ Left Click" on the {{ScreenElement|Advanced}} button of the device driver configuration. The hidden setting is found under the "[U2004A]" section | |||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] | ||
[[Category:RadiMation]] | [[Category:RadiMation]] |
Latest revision as of 10:27, 19 April 2023
A time out occurred while waiting for the measurement result
A time out occurred while waiting on a power measurement result of the U2004A. Depending on some specific settings, it can take longer than expected to perform a single power measurement. The single measurement is then taking more time than the time out, which causes this error.
Solution
Improve the measurement settings of the power meter in such a way that the measurement is faster.
- The measurement speed is related to the selected filter type. Use a lower filter setting to increase the measurement speed. The filter settings can be found in the advanced settings of the device driver under the Settings-tab. For more details see the driver configuration of the Agilent Technologies U2000A
- Increase the hidden MaxWaitTimeInSeconds property found in the 'Advanced' settings of the device driver. This hidden setting can be accessed by pressing "shift+ Left Click" on the Advanced button of the device driver configuration. The hidden setting is found under the "[U2004A]" section