− | RadiMation is searching for field sensor but no sensor in the test site, why is this happening?
| + | #REDIRECT [[FAQ#Why is an immunity test trying to control a field sensor, if no field sensor is selected in the used testsite?]] |
− | | |
− | =How to fix the problem=
| |
− | | |
− | In the configuration window of the test goto:
| |
− | {{menu|Inputs|Field probes}}
| |
− | Deselect all the field sensors in this window.
| |
− | | |
− | If the check boxes are greyed out, close all windows and add the needed [[field sensor]] to the website.
| |
− | If you do not have any add the [[virtual field sensor]] to the [[test equipment]].
| |
− | | |
− | =How does it work?=
| |
− | | |
− | In the test configuration you can specify that you want the field sensor to be measured, even if this is not the device RadiMation is levelling on.
| |
− | | |
− | For example:
| |
− | *Do a calibration on a room.
| |
− | *Put the [[EUT]] in the room
| |
− | *You put the [[Field Sensor]] on top of the [[EUT]]
| |
− | | |
− | In this example you are measuring correctly, but you just want know how much field you are generating now that the [[EUT]] is in the room.
| |
− | | |
− | You do this by telling RadiMation in the menu {{menu|Inputs|Field probes}}
| |
− | | |
− | It then know that you want the use this equipment.
| |
− | | |
− | =Why does it remove it when not configured?=
| |
− | | |
− | | |