Current sensor transfer impedance determination method: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
No edit summary
 
(2 intermediate revisions by one other user not shown)
Line 36: Line 36:
== Verification ==
== Verification ==


[[Image:Current sensor Transfer impedance Prove.PNG]]
[[Image:Current sensor Transfer impedance example.png]]


The left impedance is the signal generator which is generating enough power for 1 ampere.  
The left impedance is the signal generator which is generating enough power for 1 ampere.  
Line 48: Line 48:
So 2 ohm: <math>10*^{10}log(\frac{0,08}{50}) \approx -27,96 dB</math>
So 2 ohm: <math>10*^{10}log(\frac{0,08}{50}) \approx -27,96 dB</math>


The difference is <math>-27,96 - (-33,98) = 6,02 dB</math>
The difference to a 1 Ohm impedance is <math>-27,96 - (-33,98) = 6,02 dB</math>


==Conclusion==
==Conclusion==
Line 56: Line 56:


{{note|This method is not a replacement for a real calibration as it may be performed by a none traceable device}}
{{note|This method is not a replacement for a real calibration as it may be performed by a none traceable device}}
[[Category:RadiMation]]

Latest revision as of 07:47, 31 July 2014

Theory[edit]

Current sensor Transfer impedance Prove.PNG

The left impedance is the signal generator which is generating enough power for 1 ampere.

This 1Amp. generates in the right impedance.

The current sensor has 1 ohm transfer impedance, this means 1 ampere generates 1 Volt on the measuring part below.

The power in the lower 50 ohm impedance is

So 1 ohm:

Reference measurement[edit]

Current sensor Transfer impedance Ref measurement.PNG

Probe measurement[edit]

Current sensor Transfer impedance final measurement.PNG

Calculation[edit]

and in dBm.


Example[edit]

On 10 MHz we have the following information:

  • Calibration: 0 dBm.
  • Measurement: -27,96 dBm.

So:

Verification[edit]

Current sensor Transfer impedance example.png

The left impedance is the signal generator which is generating enough power for 1 ampere.

This 1 Ampere generates in the right impedance.

The current sensor has 2 ohm transfer impedance, this means 1 ampere generates 2 Volt on the measuring part below.

The power in the lower 50 ohm impedance is

So 2 ohm:

The difference to a 1 Ohm impedance is

Conclusion[edit]

Correction to a 1 ohm impedance is


Information.png
Note: This method is not a replacement for a real calibration as it may be performed by a none traceable device