RadiMation Error: 536881993: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
(New page: = The selected frequency step is not supported in emission tests = RadiMation has detected that unsupported type of frequency step is used in an emission test. The multiband emission ...)
 
No edit summary
 
Line 6: Line 6:
== Solution ==
== Solution ==
* Change the selected step frequency type to a linear step frequency or a logarithmic step frequency.
* Change the selected step frequency type to a linear step frequency or a logarithmic step frequency.
* If a spectrum analyzer is used, ensure that a linear step frequency is used.
* If a spectrum analyser is used, ensure that a linear step frequency is used.


[[Category:RadiMation Error]]
[[Category:RadiMation Error]]
[[Category:RadiMation]]
[[Category:RadiMation]]

Latest revision as of 08:57, 3 January 2020

The selected frequency step is not supported in emission tests[edit]

RadiMation has detected that unsupported type of frequency step is used in an emission test. The multiband emission tests, only support a linear step frequency and a logarithmic step frequency. Other step types that are available in multiband immunity tests, cannot be used in emission tests. The logarithmic step frequency is also only available if a (scanning) receiver is used. The logarithmic step frequency cannot be used in combination with a spectrum analyser.

Solution[edit]

  • Change the selected step frequency type to a linear step frequency or a logarithmic step frequency.
  • If a spectrum analyser is used, ensure that a linear step frequency is used.