RadiMation Error: 805315270: Difference between revisions
No edit summary |
No edit summary |
||
(2 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
= Detector changed due to physical constraints of the Agilent Signal | = Detector changed due to physical constraints of the Agilent Signal Analyser = | ||
This is an error message that is specific for the [[Agilent Technologies]] N90x0A series of [[Spectrum | This is an error message that is specific for the [[Agilent Technologies]] N90x0A series of [[Spectrum Analyser]]s. | ||
The Programming manual of the N90x0A series contains: | The Programming manual of the N90x0A series contains: | ||
{{Quote| | {{Quote|text= | ||
<b>Multiple Detectors</b> | <b>Multiple Detectors</b> | ||
The | The analyser always provides the requested detector on the specified trace. Depending on the detectors | ||
requested the | requested the analyser can provide up to three different detectors simultaneously within the constraints | ||
of its digital processing algorithms. Some detectors utilize more resources; the Quasi-Peak detector, for | of its digital processing algorithms. Some detectors utilize more resources; the Quasi-Peak detector, for | ||
example, utilizes most of the digital IF’s resources, and the hardware in some | example, utilizes most of the digital IF’s resources, and the hardware in some analysers is incapable of | ||
providing another detector when Quasi-Peak is on. If the limit of system resources is exceeded, | providing another detector when Quasi-Peak is on. If the limit of system resources is exceeded, | ||
detectors on some existing traces may be forced to change. When this happens, they change to match the | detectors on some existing traces may be forced to change. When this happens, they change to match the | ||
detector just requested, and a message is generated: "Detector changed due to physical | detector just requested, and a message is generated: "Detector changed due to physical | ||
constraints | constraints | ||
Example: User has traces 1, 2, and 3 with Peak, Average, and Negative Peak. User specifies QPD for | Example: User has traces 1, 2, and 3 with Peak, Average, and Negative Peak. User specifies QPD for | ||
trace 1. Traces 2 and 3 also change to QPD and we generate the message “Detector changed due to | trace 1. Traces 2 and 3 also change to QPD and we generate the message “Detector changed due to |
Latest revision as of 13:39, 9 June 2023
Detector changed due to physical constraints of the Agilent Signal Analyser
This is an error message that is specific for the Agilent Technologies N90x0A series of Spectrum Analysers.
The Programming manual of the N90x0A series contains:
“ | Multiple Detectors
The analyser always provides the requested detector on the specified trace. Depending on the detectors requested the analyser can provide up to three different detectors simultaneously within the constraints of its digital processing algorithms. Some detectors utilize more resources; the Quasi-Peak detector, for example, utilizes most of the digital IF’s resources, and the hardware in some analysers is incapable of providing another detector when Quasi-Peak is on. If the limit of system resources is exceeded, detectors on some existing traces may be forced to change. When this happens, they change to match the detector just requested, and a message is generated: "Detector changed due to physical constraints Example: User has traces 1, 2, and 3 with Peak, Average, and Negative Peak. User specifies QPD for trace 1. Traces 2 and 3 also change to QPD and we generate the message “Detector changed due to physical constraints”. Now all three traces have the QPD. |
” |
Solution
- Select a valid combination of detectors that is allowed by the N90x0A series.