<Device driver>: FTDI D2xx: IO Error[edit]

The mentioned device driver is using the FTDI D2xx DLL to communicate with the measurement device (probably over USB). However an Input/Output (IO) error has been detected.

Solution[edit]

  • Retry to perform the same operation, if this doesn't work:
    1. Stop RadiMation®
    2. Disconnect the USB cable of the mentioned measurement device
    3. Connect the USB cable of the mentioned measurement device
    4. Start RadiMation® and retry the same operation again.