RadiMation Error: 805329921: Difference between revisions
Jump to navigation
Jump to search
(One intermediate revision by the same user not shown) | |||
Line 1: | Line 1: | ||
= Amplifier Interlock | = Amplifier returned that the Group Interlock [number] is not allocated to the currently active RF path and is not evaluated. = | ||
It seems that the amplifier Interlock query returned | It seems that the amplifier Interlock query returned that the RF loop is not allocated to the currently active RF path and is not evaluated. | ||
== Solution == | == Solution == | ||
* | * Check that the correct interlock number is configured in the device settings. | ||
* Verify if this interlock group number is also used for the current RF Path. Only this specific interlock is checked by the Amplifier. | |||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] | ||
[[Category:RadiMation]] | [[Category:RadiMation]] |
Latest revision as of 15:54, 5 January 2015
Amplifier returned that the Group Interlock [number] is not allocated to the currently active RF path and is not evaluated.[edit]
It seems that the amplifier Interlock query returned that the RF loop is not allocated to the currently active RF path and is not evaluated.
Solution[edit]
- Check that the correct interlock number is configured in the device settings.
- Verify if this interlock group number is also used for the current RF Path. Only this specific interlock is checked by the Amplifier.