RadiMation Error: 805306934: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
No edit summary
No edit summary
 
(One intermediate revision by the same user not shown)
Line 4: Line 4:
== Solution ==
== Solution ==
* Restart your computer and disable that the computer can go into Hibernate or Suspend mode to prevent this error in the future.
* Restart your computer and disable that the computer can go into Hibernate or Suspend mode to prevent this error in the future.
* Also ensure that National Instruments 488.2 version 15.0 or higher is installed and used. (See: [[RadiMation Error: 805306938]])


[[Category:RadiMation Error]]
[[Category:RadiMation Error]]
[[Category:RadiMation]]
[[Category:RadiMation]]

Latest revision as of 19:39, 15 September 2017

GPIB: Hibernated NI drivers error. Restart your PC (GPIB: EPWR)[edit]

This problem is very probably caused by the effect that the computer has been changed to the Hibernate or Suspend mode. Then the National Instruments GPIB library will generate a GPIB EPWR error, which results in this error message in RadiMation.

Solution[edit]

  • Restart your computer and disable that the computer can go into Hibernate or Suspend mode to prevent this error in the future.
  • Also ensure that National Instruments 488.2 version 15.0 or higher is installed and used. (See: RadiMation Error: 805306938)