= RBW Filter settings =
= RBW Filter settings =
The ESIB has both 3 dB and 6 dB specific RBW filter settings.  Some standards specify the use of 6dB RBW filters.   
The ESIB has both 3 dB and 6 dB specific RBW filter settings.  Some standards (like CISPR16) specify the use of 6dB RBW filters.   
If the ESIB is used in (scanning) receiver mode, the 6 dB RBW filters will always be used, because no 3 dB filters are implemented in the (scanning) receiver mode.
If the ESIB is used in (scanning) receiver mode, the 6 dB RBW filters will always be used, because no 3 dB filters are implemented in the (scanning) receiver mode.
As far as we know the 6 dB RBW filters in analyzer mode of the Rohde & Schwarz ESIB are only available in firmware version 2.07 and higher. If RadiMation detects that a firmware version of 2.07 and higher is used in the ESIB it will select the 6 dB filters in analyzer mode, otherwise it will use the 3 dB filters in analyzer mode.
According to R&S release notes, the 6 dB RBW filters in analyzer mode of the Rohde & Schwarz ESIB are only available in firmware version 2.07 and higher.  
RadiMation detects which firmware version is installed in the receiver, and, if the firmware version is 2.07 or higher, the ESIB will use the 6 dB filters in analyzer mode. ESIB receivers with a firmware version below 2.07 will use the 3 dB filters in analyzer mode.


= Documentation =
= Documentation =

Revision as of 08:15, 23 May 2007

The ESIB is a full compliant Spectrum analyzer/Receiver of Rohde & Schwarz.

The ESIB was formerly known as the ESI. Several products of Rohde & Schwarz are based on the ESIB receiver or are very similar:

The ESIB developement has come to a halt, the ESIB is replaced by the Rohde & Schwarz ESU

RBW Filter settings[edit]

The ESIB has both 3 dB and 6 dB specific RBW filter settings. Some standards (like CISPR16) specify the use of 6dB RBW filters. If the ESIB is used in (scanning) receiver mode, the 6 dB RBW filters will always be used, because no 3 dB filters are implemented in the (scanning) receiver mode. According to R&S release notes, the 6 dB RBW filters in analyzer mode of the Rohde & Schwarz ESIB are only available in firmware version 2.07 and higher. RadiMation detects which firmware version is installed in the receiver, and, if the firmware version is 2.07 or higher, the ESIB will use the 6 dB filters in analyzer mode. ESIB receivers with a firmware version below 2.07 will use the 3 dB filters in analyzer mode.

Documentation[edit]

The programmer manual can be found here

Version history[edit]

  • Firmware version 4.01 was released on 2001, juni 29
  • Firmware version 4.33 was the latest version available in 2006, Februar