RadiMation Error: 805329869: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
(Created page with '= <DEVICE NAME>: Amplifier has not correctly set the desired RF Path. = This error is raised when RadiMation has send the RF Path setting to the device, received confirmation of …')
 
 
Line 1: Line 1:
= <DEVICE NAME>: Amplifier has not correctly set the desired RF Path. =
= <DEVICE NAME>: Amplifier has not correctly set the desired RF Path. =
This error is raised when RadiMation has send the RF Path setting to the device, received confirmation of the device that it has been set, RadiMation requested the current setting as final confirmation check and the device returned a different setting.
This error is raised when RadiMation has send the RF Path setting to the device, received confirmation of the device that it has been correctly received, RadiMation requested the current setting as final confirmation check and the device returned a different setting.


== Solution ==
== Solution ==

Latest revision as of 11:02, 22 July 2014

<DEVICE NAME>: Amplifier has not correctly set the desired RF Path.[edit]

This error is raised when RadiMation has send the RF Path setting to the device, received confirmation of the device that it has been correctly received, RadiMation requested the current setting as final confirmation check and the device returned a different setting.

Solution[edit]

  • Restart the test or contact your manufacture about the fact that the device is providing a command confirmation which is not correctly recalled by the device itself.