= Quasi Peak detector can only be activated when an EMI RBW bandwidth is used. =
= Quasi Peak detector can only be activated when an EMI RBW bandwidth is used. =


The Siglent SSA 3000-series only can use the Quasi peak detector, when it is the only activated detector. We detected that multiple detectors are activated, of which one is the Quasi Peak detector.
The Siglent SSA 3000-series only can use the Quasi peak detector, when the EMI RBW filter has been selected. This error is raised when the Gaussian RBW filter is selected.


== Solution ==
== Solution ==
* Change the configuration of the to be measured traces, to only have the Quasi Peak detector activated.
* Change the configuration of test to use a RBW of 200 Hz, 9 kHz or 120 kHz. The driver will automatically select the EMI RBW filter.


[[Category:RadiMation]]
[[Category:RadiMation]]
[[Category:RadiMation Error]]
[[Category:RadiMation Error]]

Latest revision as of 12:22, 16 November 2017

Quasi Peak detector can only be activated when an EMI RBW bandwidth is used.[edit]

The Siglent SSA 3000-series only can use the Quasi peak detector, when the EMI RBW filter has been selected. This error is raised when the Gaussian RBW filter is selected.

Solution[edit]

  • Change the configuration of test to use a RBW of 200 Hz, 9 kHz or 120 kHz. The driver will automatically select the EMI RBW filter.