RadiMation Error: 805339670: Difference between revisions
Jump to navigation
Jump to search
(Created page with '= Quasi Peak detector can only be activated when an EMI RBW bandwidth is used. = The Siglent SSA 3000-series only can use the Quasi peak detector, when it is the only activated …') |
No edit summary |
||
Line 1: | Line 1: | ||
= Quasi Peak detector can only be activated when an EMI RBW bandwidth is used. = | = Quasi Peak detector can only be activated when an EMI RBW bandwidth is used. = | ||
The Siglent SSA 3000-series only can use the Quasi peak detector, when | The Siglent SSA 3000-series only can use the Quasi peak detector, when the EMI RBW filter has been selected. This error is raised when the Gaussian RBW filter is selected. | ||
== Solution == | == Solution == | ||
* Change the configuration of | * Change the configuration of test to use a RBW of 200 Hz, 9 kHz or 120 kHz. The driver will automatically select the EMI RBW filter. | ||
[[Category:RadiMation]] | [[Category:RadiMation]] | ||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] |
Latest revision as of 12:22, 16 November 2017
Quasi Peak detector can only be activated when an EMI RBW bandwidth is used.[edit]
The Siglent SSA 3000-series only can use the Quasi peak detector, when the EMI RBW filter has been selected. This error is raised when the Gaussian RBW filter is selected.
Solution[edit]
- Change the configuration of test to use a RBW of 200 Hz, 9 kHz or 120 kHz. The driver will automatically select the EMI RBW filter.