Revision as of 13:05, 10 April 2019
Ticket[edit]
Also see: #9486:
Pictures[edit]
For the pictures of the schematic setup, see the powerpoint file in: [svn-trunk]:/Documentatie\TestAndMeasurementEquipment\HV-LV AttenuationCouplingSetups.pptx"
Pictures of calibration setup:
signal generator - connector - connector - analyser as powermeter
Picture of eut setup:
Signal generator - connector - HV adapter - EUT - LV Adapter - connector - analyser as powermeter
Required levels[edit]
BMW[edit]
- Signal generator level: minimum 0 dBm
- Analyser: 120 kHz RBW
- Frequency step: logarithmic
- Frequency step width: max 500 kHz
- Freq-range: 0.10 MHz - 1000 MHz
CISPR25[edit]
- Signal generator level: recommended: 0 dBm
- Minimum number of points: 401 points
- Frequency step: logarithmic