(Created page with "= There is no average probe selected = This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a RI...")
 
Line 5: Line 5:


== Solution ==
== Solution ==
Re-open your Radiated Immunity calibration test file and go to ('<Field Probe>')and make sure that at least one of the probes has the averaging checkbox checked.  
Re-open your Radiated Immunity calibration test file and go to 'Field Probe' and make sure that at least one of the probes has the averaging checkbox checked.  


[[Category:RadiMation Error]]
[[Category:RadiMation Error]]
[[Category:RadiMation]]
[[Category:RadiMation]]

Revision as of 12:34, 6 June 2019

There is no average probe selected[edit]

This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a RI calibration at least one averaging probe is mandatory to do the calibration.

Solution[edit]

Re-open your Radiated Immunity calibration test file and go to 'Field Probe' and make sure that at least one of the probes has the averaging checkbox checked.