Anritsu MS2601A: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
(Updated DeviceDriverInfo template)
(Updated DeviceDriverInfo template)
Line 2: Line 2:
|DLLFile=ADVSA.DLL
|DLLFile=ADVSA.DLL
|DeviceBrand=Anritsu
|DeviceBrand=Anritsu
|DeviceDriverName=MS2601A_PM
|DeviceDriverName=MS2601A
|DeviceType=MS2601A
|DeviceType=MS2601A
|SupportedDeviceTypes=8
|SupportedDeviceTypes=512
}}
}}



Revision as of 11:10, 10 March 2020

The Anritsu MS2601A device driver is a Spectrum Analyser which is supported by RadiMation®.  

Configuration[edit]

The following tabs are available in the advanced configuration of the Anritsu MS2601A:

ConfigurableFrequencyRangeTab.png

The frequency range of the Anritsu MS2601A as provided by the manufacturer is shown and selected as default. It is possible to overrule these frequencies and to manual adjust the allowed frequency range of the Anritsu MS2601A.

ScreenElementDescription.svg Use default start frequency If the checkbox is checked, the default start frequency will be used as the lowest usable frequency in a test for this device.
ScreenElementDescription.svg Custom start frequency If the Use default start frequency checkbox is unchecked, another start frequency (expressed in MHz) can be specified. The customized start frequency will then be used as the lowest usable frequency in a test for this device. The customized frequency can be a limitation or an extension of the default start frequency.
ScreenElementDescription.svg Use default stop frequency If the checkbox is checked, the default stop frequency will be used as the highest usable frequency in a test for this device.
ScreenElementDescription.svg Custom stop frequency If the Use default stop frequency checkbox is unchecked, another stop frequency (expressed in MHz) can be specified. The customized stop frequency will then be used as the highest usable frequency in a test for this device. The customized frequency can be a limitation or an extension of the default stop frequency.


Specifying a different frequency range can be useful if for example:

  • A device (like a coupler, antenna, injection device, cable, etc...) is still useable (but out of specification) outside the standard suggested frequency range.
  • An external mixer is used to measure an extended frequency range.
  • An up- or down-convertor is used to shift the frequency range.
  • A newer model of a device is present that has an extended frequency range, and still uses the same remote control commands.

Be careful changing these setting as RadiMation® is no longer able to verify if the Anritsu MS2601A is used outside frequency range that is specified by the manufacturer. This may result to serious damage of your measurement device.

Links

This information about an equipment is automatically generated from the list of supported device drivers from RadiMation.
If you need more information you can contact radimation-support@raditeq.com.

Manual[edit]

The programming manual of this device can be found here

Differences with other MS2601[edit]

MS2601A MS2601K
Form Factor Benchtop Benchtop
Input Impedance 50 Ohm 75 Ohm
Minimum Frequency 9 kHz 9 kHz
Maximum Frequency 2.2 GHz 2.2 GHz
Zero Span Yes Yes
Minimum Span 1 kHz 1 kHz
Maximum Span 2200 MHz 2200 MHz
Minimum Sweep Time 50 ms 50 ms
Maximum Sweep Time 1000 s 1000 s
Minimum Resolution Bandwidth 30 Hz 30 Hz
Maximum Resolution Bandwidth 1 MHz 1 MHz
Resolution Bandwidth Steps 1/3/10 1/3/10
Minimum Video Bandwidth 1 Hz 1 Hz
Maximum Video Bandwidth 100 kHz 100 kHz
Video Bandwidth Steps 1/10/100 1/10/100
Maximum Single-Side-Band Noise -80 dBc/Hz -80 dBc/Hz
Maximum Safe AC Input 25 dBm 25 dBm
Maximum DC Input 50 V 100 V
Maximum Displayed Average Noise -120 dBm -114 dBm
Maximum Dynamic Range -75 dB -75 dB
Trigger Source External,Internal External,Internal
Trigger Modes Freerun,TTL Freerun,TTL
2nd IF Input Connector Type BNC(f) BNC(f)
Probe Power No No
External Frequency Reference 10 MHz 10 MHz
Noise Source Driver No No