RadiMation Error: 805306941: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
(Created page with "= GPIB system driver error (EDVR): Ethernet communication problem = A RadiMation device driver is trying to control a GPIB device through a GPIB board. However an error is...")
 
No edit summary
 
Line 1: Line 1:
= GPIB system driver error (EDVR): Ethernet communication problem =
= GPIB system driver error (EDVR): Ethernet communication problem =
A RadiMation device driver is trying to control a [[GPIB]] device through a GPIB board. However an error is raised by the GPIB system driver that controls that GPIB board. The driver has detected that there is an ethernet communication problem with the GPIB Ethernet device.
A RadiMation device driver is trying to control a [[GPIB]] device through a GPIB board. However an error is raised by the GPIB system driver that controls that GPIB board. The driver has detected that there is an ethernet communication problem with the GPIB Ethernet device.
== National Instruments NI-488.2 ==
This error corresponds to the National Instruments NI-488.2 error 0xxE1060075.
This error is returned when the driver is unable to communicate with a GPIB Ethernet interface during an ibfind or ibdev call for one of the following reasons:
* The IP address is incorrect
* The box is not powered on
* The box did not pass its self-diagnostic test
* In general, the box is not configured or installed correctly


== Solution ==
== Solution ==

Latest revision as of 17:51, 16 February 2021

GPIB system driver error (EDVR): Ethernet communication problem

A RadiMation device driver is trying to control a GPIB device through a GPIB board. However an error is raised by the GPIB system driver that controls that GPIB board. The driver has detected that there is an ethernet communication problem with the GPIB Ethernet device.

National Instruments NI-488.2

This error corresponds to the National Instruments NI-488.2 error 0xxE1060075.

This error is returned when the driver is unable to communicate with a GPIB Ethernet interface during an ibfind or ibdev call for one of the following reasons:

  • The IP address is incorrect
  • The box is not powered on
  • The box did not pass its self-diagnostic test
  • In general, the box is not configured or installed correctly

Solution

When a National Instruments GPIB board is used, the configuration of the GPIB board can be checked and changed in the NI-MAX software.

  • Check if the GPIB Ethernet device is still present and available in the NI-MAX software.