RadiMation Error: 805330371: Difference between revisions
Jump to navigation
Jump to search
(Created page with "= An attenuation other than 0 dB or 20 dB is configured with the preamplifier enabled.\nPlease disable the preamplifier or choose an attenuation of 0 dB or 20 dB. = The Rigol...") |
No edit summary |
||
Line 1: | Line 1: | ||
= An attenuation other than 0 dB or 20 dB is configured with the preamplifier enabled. | = An attenuation other than 0 dB or 20 dB is configured with the preamplifier enabled. Please disable the preamplifier or choose an attenuation of 0 dB or 20 dB. = | ||
The Rigol Analyser only allows that an attenuation of 0 dB or 20 dB is selected when the preamplifier is activated. It is now detected that another fixed attenuator value is selected in combination with an activated preamplifier. | The Rigol Analyser only allows that an attenuation of 0 dB or 20 dB is selected when the preamplifier is activated. It is now detected that another fixed attenuator value is selected in combination with an activated preamplifier. | ||
Latest revision as of 11:15, 10 December 2021
An attenuation other than 0 dB or 20 dB is configured with the preamplifier enabled. Please disable the preamplifier or choose an attenuation of 0 dB or 20 dB.[edit]
The Rigol Analyser only allows that an attenuation of 0 dB or 20 dB is selected when the preamplifier is activated. It is now detected that another fixed attenuator value is selected in combination with an activated preamplifier.
Solution[edit]
- Modify the test configuration to disable the preamplifier.
- Modify the test configuration to use one of the allowed attenuation settings: auto attenuator, 0 dB attenuator or 20 dB attenuator.