RadiMation Error: 805310408: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 1: | Line 1: | ||
= This multislot RadiCentre requires a | = This multislot RadiCentre requires a firmware update before the firmware of a device can be updated = | ||
The firmware version of the RadiCentre is too old, to update the firmware of one of the devices in the RadiCentre. The firmware of the RadiCentre has to be updated, before the firmware of one of the devices can be updated. | The firmware version of the RadiCentre is too old, to update the firmware of one of the devices in the RadiCentre. The firmware of the RadiCentre has to be updated, before the firmware of one of the devices can be updated. At least RadiCentre firmware version 4.3.4 or higher is required. | ||
== Solution == | == Solution == | ||
* First update the firmware of the RadiCentre, after that repeat the firmware update of the device. | * First update the firmware of the RadiCentre, after that repeat the firmware update of the device. | ||
* Plug the in a RadiCentre 1, and perform the update of the device through the RadiCentre 1. | * Plug the device in a RadiCentre 1, and perform the update of the device through the RadiCentre 1. | ||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] | ||
[[Category:RadiMation]] | [[Category:RadiMation]] |
Revision as of 13:50, 24 October 2023
This multislot RadiCentre requires a firmware update before the firmware of a device can be updated
The firmware version of the RadiCentre is too old, to update the firmware of one of the devices in the RadiCentre. The firmware of the RadiCentre has to be updated, before the firmware of one of the devices can be updated. At least RadiCentre firmware version 4.3.4 or higher is required.
Solution
- First update the firmware of the RadiCentre, after that repeat the firmware update of the device.
- Plug the device in a RadiCentre 1, and perform the update of the device through the RadiCentre 1.