User:Jewe: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
Line 36: Line 36:
==Normal operation mode==
==Normal operation mode==


* Lowest frequency to 150 KHz, RBW 1 KHz.
* Lowest frequency to 150 KHz, RBW 9 KHz.
* 150 KHz to 30 MHZ with RBW 9 KHz.
* 150 KHz to 30 MHZ with RBW 1 KHz.
* 150 KHz to 30 MHZ with RBW 120 KHz.
* 30 MHz to 1000 MHz with RBW 120 KHz.
* 30 MHz to 1000 MHz with RBW 120 KHz.
* 30 MHz to 1000 MHz with RBW 9 KHz.
* 30 MHz to 1000 MHz with RBW 9 KHz.


{{note|.}}
 
{{note|All settings should be done with all the options of the machine, so Receiver/Analyser and FFT mode.}}


==Error operation mode==
==Error operation mode==

Revision as of 08:43, 31 August 2009

  1. make new getting started



Signal generator driver test[edit]

The following settings need to be tested:

Normal operation mode

  1. Amplitude
  2. Frequency
  3. AM, 80% 1 kHz
  4. FM
  5. PM 50% 1 KHz.
  6. AM external.
  7. FM external.
  8. PM external.
  9. Sine wave for amplitude modulation
  10. Square wave for amplitude modulation
Information.png
Note: All modulations needs to tested with a scope if the signal generator does not allow the settings to be read from the display.

Error operation mode

  1. unplug the communication cable during testing
  2. Set the AM to out of boundary settings, for example 0% or 1 MHz, this is not supported by the machine.
  3. Set the FM to out of boundary settings, for example 1 Hz or 1 MHz, this is not supported by the machine.
  4. Set the PM to out of boundary settings, for example 0,1 Hz and 1%.
Information.png
Note: With all out of boundary settings an error must be generated.


Spectrum analyser driver test[edit]

The following settings need to be tested:

Normal operation mode[edit]

  • Lowest frequency to 150 KHz, RBW 1 KHz.
  • Lowest frequency to 150 KHz, RBW 9 KHz.
  • 150 KHz to 30 MHZ with RBW 9 KHz.
  • 150 KHz to 30 MHZ with RBW 1 KHz.
  • 150 KHz to 30 MHZ with RBW 120 KHz.
  • 30 MHz to 1000 MHz with RBW 120 KHz.
  • 30 MHz to 1000 MHz with RBW 9 KHz.


Information.png
Note: All settings should be done with all the options of the machine, so Receiver/Analyser and FFT mode.

Error operation mode[edit]

Information.png
Note: With all out of boundary settings an error must be generated.

Getting started[edit]

Philosophy[edit]