== Solution ==
== Solution ==
It is necessary for this test that at least one check box is active. Also when no LISN is used during this test, in that case selecting one of the detectors will only change the label of the graph line, but will not have any influence on the performed test.
It is necessary for this test that at least one check box is active. Also when no LISN is used during this test, in that case selecting one of the detectors will only change the label of the graph line, but will not have any influence on the performed test.
[[Category:RadiMation Error]]
[[Category:RadiMation]]

Revision as of 15:17, 5 June 2014

No LISN Line is selected[edit]

This error is raised during the configuration of a Conducted Emission LISN TSF, when the 'Start Test' button is pressed and no detectors are selected for the LISN Line.

Solution[edit]

It is necessary for this test that at least one check box is active. Also when no LISN is used during this test, in that case selecting one of the detectors will only change the label of the graph line, but will not have any influence on the performed test.