RadiMation Error: 805309219: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
 
Line 9: Line 9:
== Solution ==
== Solution ==
* Perform a manually calibration before remote controlling the device.
* Perform a manually calibration before remote controlling the device.
* Disable the calibration in the Advanced Device Driver settings.
* Disable the automatic calibration function during the test in the Advanced Device Driver settings.


[[Category:RadiMation Error]]
[[Category:RadiMation Error]]
[[Category:RadiMation]]
[[Category:RadiMation]]

Latest revision as of 10:15, 5 May 2017

The calibration of the device caused a timeout: no confirmation could be retrieved within the maximum timeout[edit]

When the calibration setting of the device is enabled the driver will try to have the device perform a (short) calibration e.g. before the final measurement.

The first time the calibration is being activated this can take up to 90 seconds! This can cause (GPIB) timeouts, as the driver expects a confirmation after the calibration is finished. It is strongly advised to perform the first calibration of the day manually at the device itself. When it is not desirable that a calibration is performed please disable this function.

Solution[edit]

  • Perform a manually calibration before remote controlling the device.
  • Disable the automatic calibration function during the test in the Advanced Device Driver settings.