Talk:RadiMation Application Note 120: Difference between revisions
Jump to navigation
Jump to search
(→Ticket) |
No edit summary |
||
Line 3: | Line 3: | ||
== Pictures == | == Pictures == | ||
For the pictures of the schematic setup, see the powerpoint file in: | |||
[svn-trunk]:/Documentatie\TestAndMeasurementEquipment\HV-LV AttenuationCouplingSetups.pptx" | |||
Pictures of calibration setup: | Pictures of calibration setup: | ||
signal generator - connector - connector - analyser as powermeter | signal generator - connector - connector - analyser as powermeter |
Revision as of 13:04, 10 April 2019
Ticket[edit]
Also see: #9486:
Pictures[edit]
For the pictures of the schematic setup, see the powerpoint file in: [svn-trunk]:/Documentatie\TestAndMeasurementEquipment\HV-LV AttenuationCouplingSetups.pptx"
Pictures of calibration setup: signal generator - connector - connector - analyser as powermeter
Picture of eut setup: Signal generator - connector - HV adapter - EUT - LV Adapter - connector - analyser as powermeter
Required levels[edit]
BMW[edit]
- Signal generator level: minimum 0 dBm
- Analyser: 120 kHz RBW
- Frequency step: logarithmic
- Frequency step width: max 500 kHz
- Freq-range: 0.10 MHz - 1000 MHz
CISPR25[edit]
- Signal generator level: recommended: 0 dBm
- Minimum number of points: 401 points
- Frequency step: logarithmic