Release Notes/Solved items Version 2019.1.7: Difference between revisions
Jump to navigation
Jump to search
(Created page with "== RadiMation version 2019.1.7 == * #9888: Agilent 8753: Specific System Error: 50 , "CAUTION: TOO MANY SEGMENTS OR POINTS"". * #9998: N9322C analyzer gives BAND:AUTO ON er...") |
|||
Line 7: | Line 7: | ||
* #10164: Frequency column is not using settings of unit in RI frequency table. | * #10164: Frequency column is not using settings of unit in RI frequency table. | ||
* #10222: N9322C timeout | * #10222: N9322C timeout | ||
* #10224: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;:FREQ:STOP 6000000000 HZ | * #10224: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;:FREQ:STOP 6000000000 HZ | ||
* #10237: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;BAND 100000 | * #10237: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;BAND 100000 | ||
* #10255: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;:FREQ:STOP 6000000000 HZ | * #10255: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;:FREQ:STOP 6000000000 HZ | ||
Line 15: | Line 15: | ||
* #10301: Allow to select External Reference for R&S ZVB8 | * #10301: Allow to select External Reference for R&S ZVB8 | ||
* #10319: Rohde & Schwarz ESU 26 Input 1 (LAN): RSIB: Timeout has occurred. | * #10319: Rohde & Schwarz ESU 26 Input 1 (LAN): RSIB: Timeout has occurred. | ||
* #10325: Rohde & Schwarz ESPI 3 Scanning Receiver: Impossible to get description string for this Error! Probably an unknown RadiMation Error. Code: 805309369 (0x30000BB9) | |||
* #10332: Ability to use the R&S ESW with more than 32000 points | * #10332: Ability to use the R&S ESW with more than 32000 points |
Revision as of 09:55, 5 July 2019
RadiMation version 2019.1.7[edit]
- #9888: Agilent 8753: Specific System Error: 50 , "CAUTION: TOO MANY SEGMENTS OR POINTS"".
- #9998: N9322C analyzer gives BAND:AUTO ON error message
- #10073: -150dBm read out from PWR-6Ghz power sensor
- #10134: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;DET:TRAC1:FUNC POS;<Err>"\n)
- #10153: RE Manual Mode - N9322C error measuring above 1 GHZ - Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;BAND:AUTO ON;<Err>"\n
- #10164: Frequency column is not using settings of unit in RI frequency table.
- #10222: N9322C timeout
- #10224: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;:FREQ:STOP 6000000000 HZ
- #10237: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;BAND 100000
- #10255: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;:FREQ:STOP 6000000000 HZ
- #10294: Driver: R&S ZVB8 Network analyser
- #10298: Final Quasi Peak measurement of R&S ESRP is suddenly done in analyzer mode which takes a very long time
- #10300: R&S ZVB takes an additional unnecessary sweep during the port calibration
- #10301: Allow to select External Reference for R&S ZVB8
- #10319: Rohde & Schwarz ESU 26 Input 1 (LAN): RSIB: Timeout has occurred.
- #10325: Rohde & Schwarz ESPI 3 Scanning Receiver: Impossible to get description string for this Error! Probably an unknown RadiMation Error. Code: 805309369 (0x30000BB9)
- #10332: Ability to use the R&S ESW with more than 32000 points