How to program the RadiSense 10 Calibration table

In this application note we explain how you can program the RadiSense 10 Calibration table in RadiMation.

Programming the table

1) Create a RadiMation correction file.
CreateCorrection.png

2) Add the frequency, X-axis-correction, Y-axis-correction and Z-axis-correction columns.
AddColumns.png

3) Paste the earlier determined correction values from exel into the correction file.
Note. Multiple columns can be pasted from excel into a correction file with a single operation.
Note. If the first row in the table seen in the correction file is not overwritten, remove this row.
ResultingCorrectionFile.png

4) Store the correction file.
5) Go to the "advanced settings" of the RadiSense device driver and select "user calibration".
ProgrammingCorrectionTable.png

6) Make sure to check the checkbox “Use calibration if available” if the calibration should be used.
7) Click browse. Then find and select the earlier created correction file.
8) Click “Program calibration table”.
9) When the progress bar is at 100% the programming has been succesfully completed.