RadiMation Error: 805307458: Difference between revisions
Jump to navigation
Jump to search
m (2 revision(s)) |
No edit summary |
||
Line 9: | Line 9: | ||
* Check if the correct settings for the communication bus are specified | * Check if the correct settings for the communication bus are specified | ||
For RadiSense probes, which are powered by | For RadiSense probes, which are powered by laser light: | ||
* Check if the probe is 'activated', that the | * Check if the probe is 'activated', that the laser is turned on. | ||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] | ||
[[Category:RadiMation]] | [[Category:RadiMation]] |
Latest revision as of 07:35, 9 June 2023
RadiSense fieldprobe not connected
A RadiSense field probe that is tried to be controlled is not connected. The controlled device returned a response ':E9', which means that the fieldprobe is not connected.
Solution
- Check if the controlled device is powered on, and connected to the communication bus
- Check if fieldprobe is connected
- Check if the correct settings for the communication bus are specified
For RadiSense probes, which are powered by laser light:
- Check if the probe is 'activated', that the laser is turned on.