RadiMation Error: 805309219

From RadiWiki
Revision as of 10:15, 5 May 2017 by Sast (talk | contribs) (→‎Solution)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

The calibration of the device caused a timeout: no confirmation could be retrieved within the maximum timeout[edit]

When the calibration setting of the device is enabled the driver will try to have the device perform a (short) calibration e.g. before the final measurement.

The first time the calibration is being activated this can take up to 90 seconds! This can cause (GPIB) timeouts, as the driver expects a confirmation after the calibration is finished. It is strongly advised to perform the first calibration of the day manually at the device itself. When it is not desirable that a calibration is performed please disable this function.

Solution[edit]

  • Perform a manually calibration before remote controlling the device.
  • Disable the automatic calibration function during the test in the Advanced Device Driver settings.