RadiMation Error: 536872041

From RadiWiki
Revision as of 09:20, 3 January 2020 by Madero17 (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

The 'Calibration Method' should be 'signal generator level'[edit]

When an antenna scan calibration is performed, this can be performed with the 'signal generator method' or the 'network analyser method'.

  • Signal generator method: will control the signal generator to regulate the forward power to the desired power level, and measure with the spectrum analyser on the receiving antenna.
  • Network analyser method: will only use the network analyser to generate a power level on one port, and measure the received signal on the receiving antenna with the other port.

If a network analyser is present in the selected testsite the 'Network analyser method' will be used, otherwise the 'Signal generator method' will be used.

This error indicates that the selected 'Calibration method' in the TSF is set to something else than the 'Signal generator level' selection, in combination with the 'Network analyser method'. When the 'Network analyser method' is used, it is only allowed to use the 'Signal generator level' 'Calibration method'.

Solution[edit]

  • Select the 'Signal generator level' 'Calibration method'
  • Use the 'Signal generator method', by removing the 'Network analyser' device driver from the selected testsite.