The Rohde & Schwarz ESL 6 device driver is a Spectrum Analyser which is supported by RadiMation®.
Configuration
The following tabs are available in the advanced configuration of the Rohde & Schwarz ESL 6:
The frequency range of the Rohde & Schwarz ESL 6 as provided by the manufacturer is shown and selected as default. It is possible to overrule these frequencies and to manual adjust the allowed frequency range of the Rohde & Schwarz ESL 6.
Use default start frequency | If the checkbox is checked, the default start frequency will be used as the lowest usable frequency in a test for this device. |
Use default stop frequency | If the checkbox is checked, the default stop frequency will be used as the highest usable frequency in a test for this device. |
Specifying a different frequency range can be useful if for example:
- A device (like a coupler, antenna, injection device, cable, etc...) is still useable (but out of specification) outside the standard suggested frequency range.
- An external mixer is used to measure an extended frequency range.
- An up- or down-convertor is used to shift the frequency range.
- A newer model of a device is present that has an extended frequency range, and still uses the same remote control commands.
Be careful changing these setting as RadiMation® is no longer able to verify if the Rohde & Schwarz ESL 6 is used outside frequency range that is specified by the manufacturer. This may result to serious damage of your measurement device.
Performed Tests
The following tests has been performed on a R&S ESL6 to verify the functionality of the device driver
General
- Model: Rohde & Schwarz ESL 6
- Firmware version: 1.83 SP1
- Preamplifier: not tested, option (B22) was not included
- Tracking generator: not tested, option was not included
- Communication over LAN: Ok
- Test date: 2012 february 16th
Receiver mode
- RBW: Ok (tested: 200 Hz, 9 kHz, 100 kHz, 120 kHz, 300 kHz, 1 Mhz)
- Attenuator: 0, 5, 10, 15, 20, 25, 30, 35, 40, 45, 50 dB
- Initial scan: Peak, Peak + Average, Peak + Average + RMS, Peak + Average + RMS + QP
- Final Measurement Peak Detector: Ok
- Final Measurement QP Detector: Ok
- Final Measurement AVG Detector: Ok
- Final Measurement RMS Detector: Ok
- Linear frequency mode: Ok
- Logarithmic frequency mode: Ok
- Measure time: Ok
Analyser mode
- RBW: Ok (tested: 200 Hz, 9 kHz, 10 kHz, 120 kHz, 300 kHz, 1 Mhz)
- Reference Level: Ok
- Attenuator: 0, 5, 10, 15, 20, 25, 30, 35, 40, 45, 50 dB
- Sweeptime: Ok
- VBW: Ok
- Initial scan: Peak, Peak + Average, Peak + Average + RMS, Peak + Average + RMS + QP
- Point count: 101 to 32001
- Final Measurement Peak Detector: Ok
- Final Measurement QP Detector: Ok
- Final Measurement AVG Detector: Ok
- Final Measurement RMS Detector: Ok