RadiMation Error: 805309372

From RadiWiki
Revision as of 08:24, 9 April 2021 by Joro (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

Measuring at 150 kHz in a sweep with an activated "preselector" is not possible. Create a band below 150 kHz and a band above 150 kHz.

The R&S ESI analyser won't allow to set a start frequency below 150 kHz and end frequency at 150 kHz or higher with the preselector enabled.

Solution

  • Create a multiband test with a band to max 149 kHz and one band from 150 kHz and higher.
  • When not using a multiband test, then create two tests. Where the first test stops at 150 kHz, and the second test starts at 150 kHz.